Error During
Mechanical Movement
The
following messages will be displayed on the screen when error occurs. There may be
more than one possible cause and correction procedure, the priority is listed in sequence.
"****
Press switch did not release ****"
Cause: The top contact switch plate does not
release during OPEN jog cycle.
Correction:
1.) Make sure the plate is floating freely.
2.) Adjust the strain gauge PCB to 6 volts +
/ - 0.5
"****WARNING:
TOP CONTACT SWITCH NOT ENGAGE ****"
Cause: Top contact switches do not engage before
high torque press motor started.
Correction:
1.) From the Main Menu,
select TD then SW to enter the Switch Test function. Compress top contact switches
to ensure the proper response.
2.) Replace Strain Gauge
if needed.
"****TOP
CONTACT SWITCH FAIL****"
Cause: Top contact switches do not engage during
align fixture.
Correction:
1.) From the Main Menu,
select TD then SW to enter the Switch Test function. Compress top contact switches
to ensure the proper response.
2.) Replace Strain Gauge
if needed.
3.) Examine the Press
Interrupter B at the end of center shaft. Moving press up and down by tuning the
belt, the "Press interrupt B" should be toggled.
"****FRONT
OPTO NOT IN PLACE****"
Cause: The front shuttle interrupter does not
engage when shuttle reaches front.
Correction:
1.) Examine shuttle front
opto assembly.
2.) Examine data ribbon
cable for damage.
"****
REAR OPTO NOT IN PLACE****"
Cause: The rear shuttle interrupter does not
engage when shuttle reaches the back.
Correction:
1.) Examine
shuttle rear opto assembly.
2.) Examine
data ribbon cable for damage.
|